DC Voltage Measurement:
Range Resolution Accuracy / 1 Year Measurement Range
0 / 10V 1 mV 0.015% R + 2 mV -2V / + 12V
25V 1 mV 0.015% R + 2 mV -2 / + 25V
50V 1 mV 0.015% R + 2 mV -5V / + 50V
DC Current Measurement:
Range Resolution Accuracy / 1 Year Measurement Range
0 / 20mA 1 μA 0.015% R + 2 μA -6 mA / 24 mA
4 / 20mA 1 μA 0.015% R + 2 μA 3.2 mA / 24 mA
25mA 1 μA 0.015% R + 2 μA -6 mA / 25 mA
DC Voltage Source:
Range Resolution Accuracy / 1 Year Measurement Range
0 / 10V 1 mV 0.015% R + 2mV 0V / + 12V
15V 1 mV 0.015% R + 2mV 0V / + 15V
DC Current Source:
Range Resolution Accuracy / 1 Year Measurement Range
0 / 20mA 1 μA 0.015% R + 2 μA 5 00μA / 24mA
4 / 20mA 1 μA 0.015% R + 2 μA 3.2mA / 24mA
25mA 1 μA 0.015% R + 2 μA 500μA / 25mA